By examining reaction times we can see that the E-Prime recorded times far longer than
they actually were in reality. The absolute error between the target reaction time, and that
recorded by E-Prime, was 58.81 milliseconds with a standard deviation of 6.21
milliseconds. So rather than being 300 milliseconds as intended they were 359
milliseconds on average with a Standard Deviation of 6.21 milliseconds. Again these RT
errors can be easily plotted in Excel.
Response Time Error (OEM Mouse)
0.00
10.00
20.00
30.00
40.00
50.00
60.00
70.00
80.00
90.00
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Trial No
Error in ms
(M=58.81, SD=6.21, Min=46.00, Max=74.00, Var=38.56)
In this sample case study we have compared timing information obtained or generated
using the Black Box Toolkit with data recorded by the paradigm under test. From this we
can conclude that if we were to use the OEM mouse in question as a response device this
would have an unacceptable affect on our presentation and response timing. In terms of
stimulus presentation this may substantially alter what we intended when designing the
paradigm.